4.3 Article

Automated Multichannel Fluorescent Whole Slide Imaging and its Application for Cytometry

期刊

CYTOMETRY PART A
卷 75A, 期 12, 页码 1020-1030

出版社

WILEY-BLACKWELL
DOI: 10.1002/cyto.a.20791

关键词

image cytometry; slide-based cytometry; whole slide imaging; virtual microscopy; digital slide

向作者/读者索取更多资源

Slide-based image cytometry (SBC) has several advantages over flow cytometry but it is not widely used because of its low throughput, complicated workflow, and high price. Fully automated microscopes became affordable with the advent of whole slide imaging (WSI) and they can be transformed into a cytometer. A MIRAX MIDI automated whole slide imager was used with metal-halide and light emitting diode (LED)-based fluorescent illumination, filter block changer, and a cooled monochrome charge coupled device camera. The MIRAX control software was further developed for fluorescent sample detection, autofocusing, multichannel digitization, and signal correction due to nonuniform illumination. Fluorescent calibration beads were used to verify the linearity of the system. The HistoQuant software package of the MIRAX viewer was used for image segmentation and quantitative analysis. The data was displayed by the histogram, scatter plot, and gallery functions of the same program. Fluorescent samples can be reliably detected, focused, and scanned. The measured integrated fluorescence showed linearity with exposure time and staining intensity. Automated fluorescent WSI with stable LED illumination and high-quality homogeneous fluorescent slides can be used conveniently for SBC. (C) 2009 International Society for Advancement of Cytometry

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据