4.4 Article

Flow injection determination of catechol based on polypyrrole-carbon nanotube-tyrosinase biocomposite detector

期刊

CURRENT APPLIED PHYSICS
卷 10, 期 1, 页码 323-328

出版社

ELSEVIER
DOI: 10.1016/j.cap.2009.06.017

关键词

Flow injection; Biosensor; Polypyrrole; Carbon nanotube; Catechol; Tyrosinase

资金

  1. The Scientific & Technological Research Council of Turkey (TUBITAK) [105Y130]

向作者/读者索取更多资源

A flow injection catechol biosensor based on tyrosinase entrapped in carbon nanotube modified polypyrrole biocomposite film on a glassy carbon surface has been developed. Amperometric response was measured as a function of concentration of catechol, at a fixed bias voltage of -50 mV at a flow rate of 1 mL/min. The proposed biosensor exhibited impressive analytical performance such as a linear range between 3 and 50 mu M, a short response time (10 s), a detection limit of 0.671 mu M and an excellent operational (with a relative standard deviation of 0.54%) and long-term stability (85% remained after 10th week). A comparison of the analytical parameters of the developed biosensor with polypyrrole/tyrosinase film electrode was performed in the study. CNT was shown to enhance the electron transfer between the electrode and enzyme and capable to carry higher bioactivity owing to its intensified surface area. (C) 2009 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据