4.7 Review

Implementation of liquid chromatography/mass spectrometry into the clinical laboratory

期刊

CLINICA CHIMICA ACTA
卷 420, 期 -, 页码 4-10

出版社

ELSEVIER
DOI: 10.1016/j.cca.2012.10.026

关键词

Liquid chromatography; Mass spectrometry; Immunoassay; Sensitivity; Specificity

向作者/读者索取更多资源

Background: For certain clinical chemistry and toxicology analytes, liquid chromatography (LC) coupled with tandem mass spectrometry (MS/MS) offers significant advantages over traditional testing by immunoassay. Methods: Published reports comparing the performance of immunoassays against LC-MS/MS were reviewed. The tested analytes include testosterone, estradiol, thyroid hormones, vitamin D, immunosuppressants, steroids for newborn screening programs, and clinical and forensic toxicology. Results: While immunoassays are widely used in the clinical laboratory, the analytical sensitivity and specificity are inferior for many of the analytes tested in routine clinical laboratories. Moreover, LC-MS/MS can be multiplexed for high testing throughput and multiple analyte detection. The disadvantages of LC-MS/MS include the absence of Food and Drug Administration (FDA) and Conformite Europeenne (CE) Mark approved tests, the high cost of analytical instrumentation, and the technical expertise needed to operate and maintain analyzers. Conclusions: The implementation of LC-MS/MS will increase in the next few years as this technology continues to improve and the advantages become more well known. (c) 2012 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据