4.8 Article

Postprocedural Aortic Regurgitation in Balloon- Expandable and Self- Expandable Transcatheter Aortic Valve Replacement Procedures Analysis of Predictors and Impact on Long- Term Mortality: Insights From the FRANCE2 Registry

期刊

CIRCULATION
卷 129, 期 13, 页码 1415-1427

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1161/CIRCULATIONAHA.113.002677

关键词

aortic valve; aortic valve insufficiency; aortic valve stenosis

资金

  1. French Ministry of Health
  2. Edwards-Life Science
  3. Medtronic

向作者/读者索取更多资源

Background Significant postprocedural aortic regurgitation (AR) is observed in 10% to 20% of cases after transcatheter aortic valve replacement (TAVR). The prognostic value and the predictors of such a complication in balloon-expandable (BE) and self-expandable (SE) TAVR remain unclear. Methods and Results TAVR was performed in 3195 consecutive patients at 34 hospitals. Postprocedural transthoracic echocardiography was performed in 2769 (92%) patients of the eligible population, and these patients constituted the study group. Median follow-up was 306 days (Q1-Q3=178-490). BE and SE devices were implanted in 67.6% (n=1872) and 32.4% (n=897). Delivery was femoral (75.3%) or nonfemoral (24.7%). A postprocedural ARgrade 2 was observed in 15.8% and was more frequent in SE (21.5%) than in BE-TAVR (13.0%, P=0.0001). Extensive multivariable analysis confirmed that the use of a SE device was one of the most powerful independent predictors of postprocedural ARgrade 2. For BE-TAVR, 8 independent predictors of postprocedural ARgrade 2 were identified including femoral delivery (P=0.04), larger aortic annulus (P=0.0004), and smaller prosthesis diameter (P=0.0001). For SE-TAVR, 2 independent predictors were identified including femoral delivery(P=0.0001). Aortic annulus and prosthesis diameter were not predictors of postprocedural AR for SE-TAVR. A postprocedural ARgrade 2, but not a postprocedural AR=grade 1, was a strong independent predictor of 1-year mortality for BE (hazard ratio=2.50; P=0.0001) and SE-TAVR (hazard ratio=2.11; P=0.0001). Although postprocedural ARgrade 2 was well tolerated in patients with ARgrade 2 at baseline (1-year mortality=7%), it was associated with a very high mortality in other subgroups: renal failure (43%), AR

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据