One-year results of the INSPIRE trial with the novel MGuard stent: Serial analysis with QCA and IVUS

期刊

出版社

WILEY-BLACKWELL
DOI: 10.1002/ccd.22980

关键词

ACS; acute coronary syndrome; NRFL; no reflow; QCA; quantitative coronary angiography; GRFT; bypass grafts coronary; IVUS; intravascular ultrasound

向作者/读者索取更多资源

Background: The newly developed balloon-expandable Mguard stent system, a combination of an ultra-thin polymer mesh sleeve attached to the external surface of a BMS, was conceived to provide embolic protection during PCI of SVG and thrombus-containig lesions. Although the acute results (<30 days) have pointed to the efficacy of this novel device, few is known about its long-term performance. Methods: The present article address the 1-year clinical results of a cohort of 30 patients enrolled in the INSPIRE trial. Inclusion critiria was de novo lesions in SVG or native vessels with angiographic evidence of instability with potential to provoke flow disturbances and/or distal embolization. The primary endpoint (incidence of MACE-composite of cardiac death, nonfatal MI, and TLR) up to 30 days of the procedure has already been published. Secondary endpoints here presented included in-stent late lumen loss (QCA), % of stent obstruction (IVUS) at 6 months and combined MACE at 1 year. QCA and IVUS were performed by independent corelabs. Results: Mean population age was 63 years with 38% of diabetics. Overall, 55% presented with ACS and 57% of lesions were located in SVG. Most lesions had complex morphology including the presence of thrombus (26%) and ulceration (20%). Distal/proximal protection devices were not used. Preprocedural QCA data showed lesion length and reference vessel diameter of 12.0 +/- 4.5 mm and 3.0 +/- 0.5 mm. The MGuard stent was successfully delivered in all cases and final TIMI-3 was achieved in 100% with no MACE up to 30 days. At 6 months, in-stent late loss and % of stent obstruction were 1.0 +/- 0.4 mm and 28.5 > 15.6%. Up to 1 year there was no case of cardiac death, two MI (one Q-wave and one non-Q-wave) and six cases of ischemia-driven TLR. Of note, there was no case of definite/probable stent thorombosis. Conclusions: In this series of patients treated with MGuard stent, the novel device showed no midterm efficacy and safety concerns. (c) 2011 Wiley Periodicals, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据