4.6 Article

Stereotactic low-voltage x-ray irradiation for age-related macular degeneration

期刊

BRITISH JOURNAL OF OPHTHALMOLOGY
卷 95, 期 2, 页码 185-188

出版社

B M J PUBLISHING GROUP
DOI: 10.1136/bjo.2009.163907

关键词

-

向作者/读者索取更多资源

The IRay stereotactic low-voltage x-ray irradiation treatment system for age-related macular degeneration consists of a low voltage x-ray tube, an eye tracking system, a robotically controlled delivery system, a coupling device to facilitate tracking and stabilisation, a graphical user interface and gating software. Low-voltage x-rays are delivered in a series of three spots to the macula in a non-invasive manner through the inferior pars plana. These beams are designed to overlap on the centre of the macula. Each beam delivers one-third of the total dose, such that the total macula dose is three times an individual beam's dose. The device is designed to run off standard domestic electrical power, and no special shielding is necessary for the room. This system has been validated in Monte Carlo simulations, human cadaver eye studies, pre-clinical animal studies and in a phase I clinical trial.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据