期刊
BIOSENSORS & BIOELECTRONICS
卷 36, 期 1, 页码 81-88出版社
ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2012.03.044
关键词
Sulfonamide antibiotics; Integrated immunosensor; Milk; Screen-printed electrodes
类别
资金
- Spanish Ministerio de Ciencia e Innovacion [CTQ2009-09351BQU]
- AVAN-SENS from the Comunidad de Madrid [S2009PPQ-1642]
- FPU from Spanish Ministry of Education
The preparation and performance of a disposable amperometric immunosensor, based on the use of a selective capture antibody and screen-printed carbon electrodes (SPCEs), for the specific detection and quantification of sulfonamide residues in milk is reported. The antibody was covalently immobilized onto a 4-aminobenzoic acid (4-ABA) film grafted on the disposable electrode, and a direct competitive immunoassay using a tracer with horseradish peroxidase (HRP) for the enzymatic labeling was performed. The amperometric response measured at -0.2 V vs the silver pseudo-reference electrode of the SPCE upon the addition of H2O2 in the presence of hydroquinone (HQ) as mediator was used as transduction signal. The developed methodology showed very low limits of detection (in the low ppb level) for 6 sulfonamide antibiotics tested in untreated milk samples, and a good selectivity against other families of antibiotics residues frequently detected in milk and dairy products. These features, together with the short analysis time (30 min), the simplicity, and easy automation and miniaturization of the required instrumentation make the developed methodology a promising alternative in the development of devices for on-site analysis. (C) 2012 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐