4.8 Article Proceedings Paper

A glucose biosensor based on Prussian blue/chitosan hybrid film

期刊

BIOSENSORS & BIOELECTRONICS
卷 24, 期 5, 页码 1527-1530

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2008.09.025

关键词

Glucose biosensor; Prussian blue; Chitosan; Electrodeposition

向作者/读者索取更多资源

Based on electrodeposition of Prussian blue (PB) and chitosan (CS) directly on gold electrode, a hybrid film of PB/CS has been prepared. PB in this film shows a good stability compared with pure PB film when it worked in neutral and weak alkalescent solution and can act as redox mediator. It provides the potential application of such film in biosensor fabrication. A glucose biosensor was fabricated by electrodepositing glucose oxidase (GOD)/CS film on this PB/CS modified electrode. The optimum experimental conditions of biosensor for the detection of glucose have been studied in detail. Under the optimal conditions, a linear dependence of the catalytic current upon glucose concentration was obtained in the range of 2 x 10(-6) to 4 x 10(-4) M with a detection limit of 3.97 x 10(-7) M. The resulting biosensor could be applied to detect the blood sugar in real samples without any pretreatment. (c) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据