4.5 Article

Application of an improved continuous parallel shaken bioreactor system for three microbial model systems

期刊

BIOPROCESS AND BIOSYSTEMS ENGINEERING
卷 31, 期 3, 页码 193-205

出版社

SPRINGER
DOI: 10.1007/s00449-007-0183-3

关键词

continuous fermentation; parallel operation; medium consumption; oxygen transfer rate; crabtree-effect; shaking parameters

向作者/读者索取更多资源

A continuous parallel shaken bioreactor system, combining the advantages of shaken bioreactors with the advantages of continuous fermentation, was specifically manufactured from quartz glass and provides a geometric accuracy of < 1 mm. Two different model systems (facultative anaerobic bacterium C. glutamicum, and Crabtree-negative yeast P. stipitis), whose growth behaviour and metabolite formation are affected by dilution rate and oxygen availability, were studied. The transition from non-oxygen to limited conditions as function of the dilution rate could precisely be predicted applying the approach described by Maier et al. (Biochem Eng J 17:155-167, 2004). In addition, the Crabtree-positive yeast S. cerevisiae was simultaneously studied in the continuous parallel shaken bioreactor system and in a conventional 1-L bioreactor, for comparison. Essentially the same results were obtained in both types of bioreactors. However, many more reading points were obtained with the parallel shaken bioreactor system in the same time at much lower consumption of culture media.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据