期刊
BIOMEDICAL CHROMATOGRAPHY
卷 27, 期 11, 页码 1481-1488出版社
WILEY
DOI: 10.1002/bmc.2946
关键词
MIP; MEPS; LC-MS; MS; local anesthetic drugs; plasma and urine samples
This study presents the use of molecularly imprinted polymer (MIP) as packing material for microextraction by packed syringe (MEPS) to achieve higher extraction selectivity. Pentycaine was used as template for MIP. Development and validation of the determination of lidocaine, ropivacaine, mepivacaine and bupivacaine in human plasma and urine samples utilizing MIP-MEPS and liquid chromatography-tandem mass spectrometry (LC-MS/MS) were carried out. The MEPS MIP-cartridge could be used for 100 extractions before it was discarded. The extraction recovery ranged from 60 to 80%. The correlation coefficients values were >0.999 for all assays using lidocaine, ropivacaine, mepivacaine and bupivacaine in the calibration range 5-2000nmol/L. The accuracy of the studied compounds, given as a percentage variation from the nominal concentration values, ranged from -4.9 to 8.4% using plasma and urine samples. The between-batch precision, given as the relative standard deviation, at three different concentrations (quality control samples) was ranged from -4.7 to 14.0% and from 1.8 to 12.7% in plasma and urine, respectively. The lower limit of quantification and limit of detection of the studied substances were 5.0 and 1.0nm, respectively. Copyright (c) 2013 John Wiley & Sons, Ltd.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐