4.6 Article

Does the Ingrowth Surface Make a Difference? A Retrieval Study of 423 Cementless Acetabular Components

期刊

JOURNAL OF ARTHROPLASTY
卷 30, 期 4, 页码 706-712

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2014.10.028

关键词

acetabular shells; bone ingrowth; implant retrieval

向作者/读者索取更多资源

The effect of factors such as design, alloy and coating type on bony or fibrous tissue ingrowth was evaluated in a study of 423 retrieved cementless acetabular shells representing 16 shell designs. Small-beaded (250 pm) porous coatings, either with or without hydroxyapatite (HA) coatings, proved to be the superior porous surface for bone ingrowth. Small-beaded shells that were Duofix coated had predominantly fibrous tissue ingrowth. In addition to bead size, alloy type and surface type have significant effect on bone ingrowth. In contrast, there is no significant association between bone ingrowth and time in situ, with most bone ingrowth occurring early. Although roughened, press-fit shells have acceptable clinical and Registry data, they showed some of the lowest ingrowth/ongrowth scores of all the shells tested. Crown Copyright (C) 2014 Published by Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据