4.4 Article

Microfluidic reactor geometries for radiolysis reduction in radiopharmaceuticals

期刊

APPLIED RADIATION AND ISOTOPES
卷 70, 期 8, 页码 1691-1697

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.apradiso.2012.03.004

关键词

Microfluidics; Radiochemistry; Autoradiolysis; Radiotracer synthesis; Lab-on-chip; PET

资金

  1. Leading-Edge Cluster m4
  2. German Federal Ministry of Education and Research

向作者/读者索取更多资源

Autoradiolysis describes the degradation of radioactively labeled compounds due to the activity of the labeled compounds themselves. It scales with activity concentration and is of importance for high activity and microfluidic PET tracer synthesis. This study shows that microfluidic devices can be shaped to reduce autoradiolysis by geometric exclusion of positron interaction. A model is developed and confirmed by demonstrating in-capillary storage of non-stabilized [F-18]FDG (2-[F-18]Fluoro-2-deoxy-D-glucose) at max. 23 GBq/ml while maintaining > 90% radiochemical purity over 14 h. (C) 2012 Elsevier Ltd. All rights reserved.