4.4 Article

An examination of mass thickness measurements with X-ray sources

期刊

APPLIED RADIATION AND ISOTOPES
卷 66, 期 10, 页码 1387-1391

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.apradiso.2008.04.012

关键词

X-ray; mass attenuation; mass thickness; equivalent energy; Monte Carlo; EGSnrc

向作者/读者索取更多资源

A method using X-rays to measure mass thickness is discussed. The method utilizes a filter to absorb low energy and characteristic photons so that the hardened X-ray spectra have a peaked energy distribution. An equivalent X-ray energy, which defines the attenuation in a material of interest can be used. The effect on the X-ray spectra of different filters is examined with Monte Carlo simulation using the EGSnrc package. A theoretical model for X-ray absorption that shows that the method can achieve good precision for a certain range of mass thicknesses is advanced. Experimental results agree well with the theoretical analysis. It is found that for a certain range of mass thicknesses, the relative error can be less than 1% for the aluminum alloy sample at the tube voltage of 30 or 45 kV. (c) 2008 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据