4.8 Article

Autonomous in Situ Analysis and Real-Time Chemical Detection Using a Backpack Miniature Mass Spectrometer: Concept, Instrumentation Development, and Performance

期刊

ANALYTICAL CHEMISTRY
卷 86, 期 6, 页码 2900-2908

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac403765x

关键词

-

资金

  1. Department of Defense [H92222-10-C0028]
  2. Department of Homeland Security [HSHQDC-09-9-00008]

向作者/读者索取更多资源

A major design objective of portable mass spectrometers is the ability to perform in situ chemical analysis on target samples in their native states in the undisturbed environment. The miniature instrument described here is fully contained in a wearable backpack (10 kg) with a geometry-independent low-temperature plasma (LTP) ion source integrated into a hand-held head unit (2 kg) to allow direct surface sampling and analysis. Detection of chemical warfare agent (CWA) simulants, illicit drugs, and explosives is demonstrated at nanogram levels directly from surfaces in near real time including those that have complex geometries, those that are heat-sensitive, and those bearing complex sample matrices. The instrument consumes an average of 65 W of power and can be operated autonomously under battery power for ca. 1.5 h, including the initial pump-down of the manifold. The maximum mass-to-charge ratio is 925 Th with mass resolution of 1-2 amu full width at half-maximun (fwhm) across the mass range. Multiple stages of tandem analysis can be performed to identify individual compounds in complex mixtures. Both positive and negative ion modes are available. A graphical user interface (GUI) is available for novice users to facilitate data acquisition and real-time spectral matching.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据