A novel label-free immunosensor for the detection of C-reactive protein (CRP) was developed based on a three-dimensional ordered macroporous (3DOM) gold film modified electrode by using the electrochemical impedance spectroscopy (EIS) technique. The electrode was electrochemically fabricated with an inverted opal template, making the surface area of the 3DOM gold film up to 14.4 times higher than that of a classical bare flat one, characterized by the cyclic voltammetric (CV) technique. The 3DOM gold film which was composed of interconnected gold nanoparticles not only has a good biocompatible microenvironment but also promotes the increase of conductivity and stability. The CRP immunosensor was developed by covalently conjugating CRP antibodies with 3-mercaptopropionic acid (MPA) on the 3DOM gold film electrode. The CRP concentration was measured through the increase of impedance values in the corresponding specific binding of CRP antigen and CRP antibody. The increased electron-transfer resistance (R-et) values were proportional to the logarithmic value of CRP concentrations in the range of 0.1 to 20 ng mL(-1). The detection of CRP levels in three sera obtained from hospital showed acceptable accuracy.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据