4.7 Article

Monolithic spin column extraction and GC-MS for the simultaneous assay of diquat, paraquat, and fenitrothion in human serum and urine

期刊

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 400, 期 1, 页码 25-31

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-010-4633-4

关键词

Diquat; Paraquat; Fenitrothion; Monolitic spin column extraction; GC-MS

资金

  1. Japan Society for the Promotion of Science [21590745]
  2. Grants-in-Aid for Scientific Research [21590745] Funding Source: KAKEN

向作者/读者索取更多资源

We present a method based on monolitic spin column extraction and gas chromatography-mass spectrometry as an analytical method for screening diquat (DQ), paraquat (PQ), and fenitrothion in serum and urine. This method is useful for clinical and forensic toxicological analyses. Recovery of DQ, PQ, and fenitrothion from serum and urine, spiked at concentrations between 0.1, 2.5, 20, and 45 mu g/ml, ranged from 51.3% to 106.1%. Relative standard deviation percentages were between 3.3% and 14.8%. Detection and quantitation limits for serum and urine were 0.025 and 0.05 mu g/ml, respectively, for DQ, 0.1 and 0.1 mu g/ml, respectively, for PQ, and 0.025 and 0.05 mu g/ml, respectively, for fenitrothion. Therefore, these compounds can be detected and quantified in the case of acute poisoning.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据