ANALYTICA CHIMICA ACTA
卷 770, 期 -, 页码 153-160出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2013.01.045
关键词
Cathodic electrochemiluminescence; Disposable biosensor; Immobilized Ru(bpy)(3)(2+); Uric acid determination
资金
- Ministerio de Ciencia e Innovacion, Direccion General de Investigacion y Gestion del Plan Nacional de I+D+i (Spain) [CTQ2009-14428-C02-01, CTQ2009-14428-C02-02]
- Junta de Andalucia (Proyecto de Excelencia) [P08-FQM-3535]
- European Regional Development Funds (ERDF)
A new method for uric acid (UA) determination based on the quenching of the cathodic ECL of the tris(2,2-bipyridine)ruthenium(II)-uricase system is described. The biosensor is based on a double-layer design containing first tris(2,2-bipyridine)ruthenium(II) (Ru(bpy)(3)(2+)) electrochemically immobilized on graphite screen-printed cells and uricase in chitosan as a second layer. The uric acid biosensing is based on the ECL quenching produced by uric acid over the cathodic ECL caused by immobilized Ru(bpy)(3)(2+) in the presence of uricase. The use of a -1.1 V pulse for 1 s with a dwelling time of 10 s makes it possible to estimate the initial enzymatic rate, which is used as the analytical signal. The Stern-Volmer type calibration function shows a dynamic range from 1.0 x 10(-5) to 1.0 x 10(-3) M with a limit of detection of 3.1 x 10(-6) M and an accuracy of 13.6% (1.0 x 10(-4) M, n=5) as relative standard deviation. Satisfactory results were obtained for urine samples, creating an affordable alternative for uric acid determination. (C) 2013 Elsevier B.V. All rights reserved.