期刊
ANALYTICA CHIMICA ACTA
卷 625, 期 2, 页码 117-130出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2008.07.024
关键词
inductively coupled plasma mass; spectrometry; major; trace; ultratrace; fusion; geochemical analysis
资金
- Spanish MEC [CGL2005-03157]
- UPV/EHU [GIU05/02]
A method has been developed for the simultaneous quantification of major to ultratrace elements in geological samples using quadrupole ICP-MS techniques. The sample preparation involves fusion with LiBO2 and dilution in HNO3-HF which allows complete decomposition of refractory minerals and quantification of the elements of interest. The effects of high Total Dissolved Solids (TDS) and Li in the solution are minimized using a matrix-tolerant interface and conditioning the instrument with LiBO2 solution. The signal drift is moreover controlled using conventional internal standards and specific Drift Correction Standards (DCS). A key issue of the technique is the external calibration using selected Certified Reference Materials (CRM). Depending on the sample type and analytes of interest three optimized programmable modes are used sequentially: Standard, Collision Cell (CCT) and Kinetic Energy Discrimination (KED) mode. The method allows to quantify more than 40 elements in concentrations from tens-of-percent to <0.1 ppm levels during a single experiment. The method has been validated through the analysis of different CRMs with recovery factors of ca. 100% and typical 2 sigma errors of <10%. (C) 2008 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据