4.6 Article

A lateral flow paper microarray for rapid allergy point of care diagnostics

期刊

ANALYST
卷 139, 期 10, 页码 2348-2354

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c3an01806g

关键词

-

资金

  1. Swedish Government Agency for Innovation Systems (VINNOVA)
  2. ProNova VINN Excellence Centre for Protein Technology

向作者/读者索取更多资源

There is a growing need for multiplexed specific IgE tests that can accurately evaluate patient sensitization profiles. However, currently available commercial tests are either single/low-plexed or require sophisticated instrumentation at considerable cost per assay. Here, we present a novel convenient lateral flow microarray-based device that employs a novel dual labelled gold nanoparticle-strategy for rapid and sensitive detection of a panel of 15 specific IgE responses in 35 clinical serum samples. Each gold nanoparticle was conjugated to an optimized ratio of HRP and anti-IgE, allowing significant enzymatic amplification to improve the sensitivity of the assay as compared to commercially available detection reagents. The mean inter-assay variability of the developed LFM assay was 12% CV, and analysis of a cohort of clinical samples (n = 35) revealed good general agreement with ImmunoCAP, yet with a varying performance among allergens (AUC = [0.54-0.88], threshold 1 kU). Due to the rapid and simple procedure, inexpensive materials and read-out by means of a consumer flatbed scanner, the presented assay may provide an interesting low-cost alternative to existing multiplexed methods when thresholds > 1 kU are acceptable.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据