4.5 Article

Design and Laboratory Evaluation of a Sequential Spot Sampler for Time-Resolved Measurement of Airborne Particle Composition

期刊

AEROSOL SCIENCE AND TECHNOLOGY
卷 48, 期 6, 页码 655-663

出版社

TAYLOR & FRANCIS INC
DOI: 10.1080/02786826.2014.911409

关键词

-

资金

  1. NIH [RC3ES019081]

向作者/读者索取更多资源

A new sampling approach has been developed to enable affordable, time-resolved monitoring of particulate chemical composition, and more generally to provide concentrated samples of airborne particles. Using a newly developed, moderated water-based condensational growth technology, individual particle samples are deposited in a 1-mm diameter dry spot. The moderated condensation technology enables this collection with minimal temperature rise, providing robust collection for volatile constituents. Measured collection efficiencies are above 95% for particles in the size range from 0.010 mu m to 2.5 mu m. A set of 20 or more time-resolved samples, plus blanks, may be collected onto a multiwell collection plate. For chemical analysis the plate is returned to the laboratory, and placed directly into a modified autosampler, without extraction or preparation. The autosampler handles the addition of eluent, extraction, and sample injection without user manipulation. This article presents the design and laboratory evaluation of a 1.5 L/min sampling rate version of this system. Copyright 2014 American Association for Aerosol Research