期刊
AEROSOL SCIENCE AND TECHNOLOGY
卷 43, 期 9, 页码 866-871出版社
TAYLOR & FRANCIS INC
DOI: 10.1080/02786820902998373
关键词
-
类别
资金
- United States National Science Foundation's [ATM-0349292, ATM-0337853]
The nephelometer, the primary instrument used for accurate in-situ scattering measurements, suffers from small errors due to truncation and angular nonidealities in the light source. Because that error depends on particle size, it has been traditionally estimated from the wavelength dependence of scattering, which also depends on particle size. Absorption by particles alters the scattering wavelength dependence, so the traditional correction is in error by 1-5% for absorbing particles, particularly when that absorption is wavelength-dependent. Single-scattering albedos under 0.9 and absorption angstrom ngstrom exponents above 2 can have errors of over 2%. This problem will occur only very near sources, including laboratory measurements of combustion aerosol. For these situations, we suggest the correction should be calculated using Mie theory and an assumed refractive index, resulting in less than a 2% error. Scattering wavelength dependence for absorbing particles should not be used as a measurement of particle size.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐