Thermal Ablation: A Comparison of Thermal Dose Required for Radiofrequency-, Microwave-, and Laser-Induced Coagulation in an Ex Vivo Bovine Liver Model

期刊

ACADEMIC RADIOLOGY
卷 16, 期 12, 页码 1539-1548

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.acra.2009.06.016

关键词

Thermal ablation; radiofrequency tumor ablation; thermal dose; image-guided intervention; hyperthermia

资金

  1. National Institute of Health (NIH) [NCI-R01EB0004-84-01A1]

向作者/读者索取更多资源

Rationale and Objectives. To compare thermal dosimetry metrics for specified diameters of coagulation achieved using three different ablation energy Sources. Materials and Methods. 204 ablations measuring 20, 30, or 40 +/- 2 mm were created in an ex-vivo bovine liver model using 1) 2.5 cm cluster RF electrodes (n = 114), 2) 3 cm microwave antennas (n = 45), and 3) 3 cm laser diffusing fibers (n = 45). Continuous temperature monitoring was performed 5-20 mm from the applicators 10 calculate: a) the area under the Curve (AUC), b) Cumulative equivalent minutes at 43 degrees C (CEM43), and c) Arrhenius damage integral (Omega) for the critical ablation margin (DOC), with results compared by multivariate analysis of variance and regression analysis. Results. The end temperatures at the margin of coagulation varied, and was lowest for the RF cluster electrode (33-58 degrees C). higher for laser (52-72 degrees C), and covered the widest range for microwave (42-95 degrees C). These end temperatures correlated with applied energy, as linear functions (r(2) = 0.74-0.96). The total beat needed to achieve ablation (AUC) varied with applied energy and Coagulation diameter as negative exponential (RF and laser) or negative power (microwave) functions (r(2) = 0.82-0.98). Similarly, CEM43 values varied exponentially with energy and distance (r(2) = 0.52-0.76) over it wide range of values (10(12)). Likewise, Omega varied not only based upon energy source and DOC, but also as a positive linear correlation to applied energy and with sigmoid correlation to duration of ablation (r(2) = 0.85-0.97). Conclusion. Our Study demonstrates that the thermal dosimetry of ablation is not based solely on it fixed end temperature at the margin of the Coagulation zone. Thermal dosimetry is not constant, but dependent on the type and amount of energy applied and distance Suggesting the need to take into account the rate of heat transfer for ablation dosimetry.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据