4.2 Article

Analysis of a Pulse Rate Variability Measurement Using a Smartphone Camera

期刊

JOURNAL OF HEALTHCARE ENGINEERING
卷 2018, 期 -, 页码 -

出版社

HINDAWI LTD
DOI: 10.1155/2018/4038034

关键词

-

资金

  1. European Union [GINOP-2.2.1-15-2017-00073]
  2. European Regional Development Fund

向作者/读者索取更多资源

Background. Heart rate variability (HRV) provides information about the activity of the autonomic nervous system. Because of the small amount of data collected, the importance of HRV has not yet been proven in clinical practice. To collect population-level data, smartphone applications leveraging photoplethysmography (PPG) and some medical knowledge could provide the means for it. Objective. To assess the capabilities of our smartphone application, we compared PPG (pulse rate variability (PRV)) with ECG (HRV). To have a baseline, we also compared the differences among ECG channels. Method. We took fifty parallel measurements using iPhone 6 at a 240 Hz sampling frequency and Cardiax PC-ECG devices. The correspondence between the PRV and HRV indices was investigated using correlation, linear regression, and Bland-Altman analysis. Results. High PPG accuracy: the deviation of PPG-ECG is comparable to that of ECG channels. Mean deviation between PPG-ECG and two ECG channels: RR: 0.01 ms-0.06 ms, SDNN: 0.78 ms-0.46 ms, RMSSD: 1.79 ms-1.21 ms, and pNN50: 2.43%-1.63%. Conclusions. Our iPhone application yielded good results on PPG-based PRV indices compared to ECG-based HRV indices and to differences among ECG channels. We plan to extend our results on the PPG-ECG correspondence with a deeper analysis of the different ECG channels.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据