4.5 Article

Self-calibration for lab-μCT using space-time regularized projection-based DVC and model reduction

期刊

MEASUREMENT SCIENCE AND TECHNOLOGY
卷 29, 期 2, 页码 -

出版社

IOP Publishing Ltd
DOI: 10.1088/1361-6501/aa9818

关键词

x-ray computed tomography; online calibration; artifact correction; digital volume correlation; proper generalized decomposition (PGD)

资金

  1. French 'Agence Nationale de la Recherche' [ANR-10-EQPX-37 MATMECA]
  2. COMINSIDE [ANR-14-CE07-0034-02]

向作者/读者索取更多资源

An online calibration procedure for x-ray lab-CT is developed using projection-based digital volume correlation. An initial reconstruction of the sample is positioned in the 3D space for every angle so that its projection matches the initial one. This procedure allows a space-time displacement field to be estimated for the scanned sample, which is regularized with (i) rigid body motions in space and (ii) modal time shape functions computed using model reduction techniques (i.e. proper generalized decomposition). The result is an accurate identification of the position of the sample adapted for each angle, which may deviate from the desired perfect rotation required for standard reconstructions. An application of this procedure to a 4D in situ mechanical test is shown. The proposed correction leads to a much improved tomographic reconstruction quality.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据