期刊
ANALYTICAL BIOCHEMISTRY
卷 545, 期 -, 页码 54-64出版社
ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.ab.2018.01.009
关键词
Voltammetry; Linezolid; Meropenem; Theophylline; Carbon nanotubes; Plasma
Treatment of healthcare associated Pneumonia (HCAP) caused by Methicillin-resistant Staphylococcus aureus (MRSA) requires therapeutic protocols formed of linezolid (LIN) either alone or in combination with meropenem (MERO) and theophylline (THEO). The inter-individual pharmacokinetic variations require the development of reliable therapeutic drug monitoring (TDM) tools especially in immunocompromised patients. A sensitive square wave voltammetric sensor using multiwalled carbon nanotubes (MWCNTs) modified carbon paste electrode in Britton-Robinson buffer was developed and validated. Experimental parameters such as pH, percentage of MWCNTs, and pre-concentration time were optimized. The sensor was employed at pH 11.0 for the determination of LIN in plasma within a concentration range of 2.5 x 10(-8) - 8.0 x 10(-6) mol L-1 without interference from co-administered medications. On the other hand, simultaneous monitoring of LIN, MERO and THEO in plasma was feasible at pH 3.0 over concentration ranges of 4.0 x 10(-7) - 9.0 x 10(-5), 8.0 x 10(-7) - 9.0 x 10(-5) and 8.0 x 10(-7) - 9.0 x 10(-5) mol L-1, respectively. The performance of the proposed sensor was validated and the applicability for TDM has been demonstrated in plasma of healthy volunteers.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据