Usefulness of Sealants for Dural Closure: Evaluation in an In Vitro Model

期刊

OPERATIVE NEUROSURGERY
卷 15, 期 4, 页码 425-431

出版社

OXFORD UNIV PRESS INC
DOI: 10.1093/ons/opx260

关键词

Cerebrospinal fluid leakage; Dura mater; Dural closure

资金

  1. Polyganics bv, Groningen, The Netherlands

向作者/读者索取更多资源

BACKGROUND: Cerebrospinal fluid (CSF) leakage occurs in 4% to 32% of cranial surgeries and is associated with significant patient burden and expense. The use of sealant as an adjunct to primary dural closure is assumed to help prevent CSF leakage. OBJECTIVE: To examine the utility of different sealants for dural closure using an in Vitro model. METHODS: We evaluated 9 commonly used dural sealants, including Tachosil (Takeda Inc, Osaka, Japan), Adherus (Hyperbranch Inc, Durham, North Carolina), Duraform (Codman, Raynham, Massachusetts), Tissudura (Baxter, Deerfield, Illinois), Hemopatch (Baxter), TissuePatchDural (Tissuemed, Leeds, United Kingdom), Tisseel (Baxter), Duragen Secure (Integra, Plainsboro, New Jersey), and Duraseal, (Integra). Sealants were tested in 2 novel in Vitro setups using fresh porcine dura: the first tested the acute burst pressure of a sealed 3-mm gap, while the second examined resistance to a pressure wave mimicking intracranial pressure for 72 h. RESULTS: Adherus showed the highest mean burst pressure (87 +/- 47 mmHg) followed by Tachosil (71 +/- 16 mmHg) and Duraseal (51 +/- 42 mmHg); these were the only 3 sealants showing burst pressures above normal physiological intracranial pressure. In the 72-h setup, only Adherus and Duraseal maintained appropriate sealing for the duration of the experiment. Tachosil released from the dura after 1.4 h (95% confidence interval, -1.8-4.7). CONCLUSION: Given the high cost of sealants and the results of this study, we advocate a critical attitude toward sealant application as an adjunct to classic dural closure.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据