4.7 Article

Simple and rapid differentiation of toxic gases using a quartz crystal microbalance sensor array coupled,with principal component analysis

期刊

SENSORS AND ACTUATORS A-PHYSICAL
卷 263, 期 -, 页码 1-7

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.sna.2017.05.039

关键词

QCM sensor array; Graphene oxide; N-substituted pyrrole derivative; Differentiation of toxic gases

向作者/读者索取更多资源

A novel quartz crystal microbalance (QCM) sensor array was fabricated for differentiating three toxic gases (NH3, CO and NO2) from their tertiary mixtures. The QCM sensor array composed of three QCM sensors that were constructed using graphene oxide (GO)-based and N-substituted pyrrole derivative based films. The response frequency changes of a QCM sensor array were recorded upon exposing to each target toxic gas with varying concentrations and the obtained data were analyzed by principal component analysis (PCA) in order to differentiate the target toxic gas under testing condition of tertiary mixing of toxic gases. The results of this work show promising prospects for the use of a low-cost, easy to use and rapid method for differentiating the target toxic gas in the presence of a mixture of gases. (C) 2017 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据