4.5 Article

An Electrochemical Biosensor with Uricase Immobilized on Functionalized Gold Coated Copper Wire Electrode for Urinary Uric Acid Assay

期刊

ELECTROANALYSIS
卷 28, 期 4, 页码 695-703

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/elan.201500539

关键词

Copper wire electrode; Electrochemical biosensor; Uric acid; Uricase; Urine

资金

  1. National Science Council [NSC 102-2112-M-033-006]

向作者/读者索取更多资源

A new second generation uricase electrode for urinary uric acid determination has been developed by chemically binding both uricase and redox mediator to inexpensive copper wire through simple electrodeposition of gold on copper surface and subsequent functionalization of the gold with L-methionine. During a 209-day testing period, the overall electrode performance exhibits in average a low oxidation potential of 0.33 V, a response time of 5s, a widest linear calibration concentration range (0-2.38 mM, r(2) > 0.9952), a sensitivity of 50 mu A mM(-1), and a detection limit of 2.4 mu M. The measurement accuracy and precision for the determination of uric acid in human urine specimens were 85.6-95.5% and 0.3-2.4%, respectively. The developed uricase electrode is potential for clinical applications.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据