4.4 Article

The effect of bone growth onto massive prostheses collars in protecting the implant from fracture

期刊

MEDICAL ENGINEERING & PHYSICS
卷 41, 期 -, 页码 19-25

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.medengphy.2016.12.007

关键词

Prostheses; Orthopaedics; Finite Element Analysis; HA collar; Bone growth

资金

  1. Orthopaedic Research UK

向作者/读者索取更多资源

Limb-sparing distal femoral endoprotheses used in cancer patients have a high risk of aseptic loosening. It had been reported that young adolescent patients have a higher rate of loosening and fatigue fracture of intramedullary stems because the implant becomes undersized as patients grow. Extracortical bone growth into the grooved hydroxyapatite-coated collar had been shown to reduce failure rates. The stresses in the implant and femur have been calculated from Finite Element models for different stages of bone growth onto the collar. For a small diameter stem without any bone growth, a large stress concentration at the implant shoulder was found, leading to a significant fracture risk under normal walking loads. Bone growth and osseointergration onto the implant collar reduced the stress level in the implant to safe levels. For small bone bridges a risk of bone fracture was observed. (C) 2016 IPEM. Published by Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据