Effect of spatial resolution and filtering on mapping cardiac fibrillation

期刊

HEART RHYTHM
卷 14, 期 4, 页码 608-615

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.hrthm.2017.01.023

关键词

Atrial fibrillation; Ventricular fibrillation; Rotor; Optical mapping; Multielectrode mapping; Electroderesolution; Signal filters

资金

  1. la Caixa Foundation, Spain
  2. Heart and Stroke Foundation Clinician Scientist

向作者/读者索取更多资源

Background Endocardial mapping tools use variable interelectrode resolution, whereas body surface mapping tools use narrow bandpass filtering (BPF) to map fibrillatory mechanisms established by high-resolution optical imaging. Objective The purpose of this study was to study the effect of resolution and BPF on the underlying mechanism being mapped. Methods Hearts from 14 healthy New Zealand white rabbits were Langendorff perfused. We studied the effect of spatial resolution and BPF on the location and characterization of rotors by comparing phase singularities detected by high-resolution unfiltered optical maps and of fibrillating myocardium with decimated and filtered maps with simulated electrode spacing of 2, 5, and 8 mm. Results As we decimated the maps with 2-mm, 5-mm, and 8-mm interelectrode spacing, the mean (+/- SD) number of rotors detected decreased from 10.2 +/- 9.6, 1.6 +/- 3.2, and 0.2 +/- 0.5, respectively. Lowering the resolution led to synthesized pseudo-rotors that may be inappropriately identified. Applying a BPF led to fewer mean phase singularities detected (248 +/- 207 vs 333 +/- 130; P <.01), giving the appearance of pseudo-spatial stability measured as translation index (with BPF 3.6 +/- 0.4 mm vs 4.0 +/- 0.5 mm without BPF; P <.01) and pseudo-temporal stability with longer duration (70.0 +/- 17.6 ms in BPF maps vs 44.1 +/- 6.6 ms in unfiltered maps; P <.001) than true underlying fibrillating myocardium mapped. Conclusion Electrode resolution and BPF of electrograms can result in distortion of the underlying electrophysiology of fibrillation. Newer mapping techniques need to demonstrate sensitivity analysis to quantify the degree of distortion before clinical use to avoid inaccurate electrophysiologic interpretation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据