4.6 Article

Computer-Assisted Optimization of Electrodeposited Hydroxyapatite Coating Parameters on Medical Alloys

出版社

SPRINGER
DOI: 10.1007/s11661-016-3341-x

关键词

-

资金

  1. Scientific and Technological Research Council of Turkey (TUBITAK)

向作者/读者索取更多资源

CoCrMo bio-metallic alloys were coated with a hydroxyapatite (HA) film by electrodeposition using various electrochemical parameters. Response surface methodology and central composite design were used to optimize deposition parameters such as electrolyte pH, deposition potential, and deposition time. The effects of the coating parameters were evaluated within the limits of solution pH (3.66 to 5.34), deposition potential (-1.13 to -1.97 V), and deposition time (6.36 to 73.64 minutes). A 5-level-3-factor experimental plan was used to determine ideal deposition parameters. Optimum conditions for the deposition parameters of the HA coating with high in vitro corrosion performance were determined as electrolyte pH of 5.00, deposition potential of -1.8 V, and deposition time of 20 minutes.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据