期刊
CHROMATOGRAPHIA
卷 -, 期 -, 页码 -出版社
SPRINGER HEIDELBERG
DOI: 10.1007/s10337-023-04286-9
关键词
HPLC-PDA; NSAIDs; Non-Steroidal Anti-Inflammatory Drugs; Reversed-phase HPLC; Tandem Columns
A tandem-column liquid chromatographic method was developed to efficiently separate ten non-steroidal anti-inflammatory drugs using a suitable tandem-column configuration.
A tandem-column liquid chromatographic method was developed to separate ten non-steroidal anti-inflammatory drugs (Aceclofenac, Aspirin, Carprofen, Celecoxib, Diclofenac, Flurbiprofen, Ibuprofen, Ketoprofen, Naproxen, Salicylic Acid) using an achiral tandem-column configuration under reversed-phase conditions. An NSAID test mix was prepared with 0.5 mg/mL thiourea as a t0 marker and 0.5 mg/mL of the above NSAIDs in 1:1 acetonitrile: 0.2% formic acid. Three different 5-cm columns were investigated individually and in two different tandem-column configurations: an octyl (C8) column, a perfluorophenyl (PFP) column, and a biphenyl (BP) column. The two tandem configurations investigated were a PFP column followed by either a BP or C8 column. While none of the individual columns provided the selectivity needed to separate two sets of critical pairs (Ketoprofen and Naproxen; Flurbiprofen and Ibuprofen), the unique selectivity provided by the two tandem-column combinations enabled both critical pairs to be separated. The PFP + BP tandem column combination was able to separate the 10 NSAIDs with baseline or greater resolution in less than 25 min.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据