4.4 Article

Experimental investigation and statistical modeling of temperature rise in rotary ultrasonic bone drilling

期刊

MEDICAL ENGINEERING & PHYSICS
卷 38, 期 11, 页码 1330-1338

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.medengphy.2016.08.012

关键词

RUD; Bone; Thermocouple; RSM; ANOVA; Temperature

资金

  1. EPSRC-DST project titled as Modelling of Advanced Materials for Simulation of Trans formative Manufacturing Process (MAST) [EP/K028316/1, DST/RC-UK/14-AM/2012]

向作者/读者索取更多资源

Thermal necrosis is one of the major problems associated with the bone drilling process in orthopedic/trauma surgical operations. To overcome this problem a new bone drilling method has been introduced recently. Studies have been carried out with rotary ultrasonic drilling (RUD) on pig bones using diamond coated abrasive hollow tools. In the present work, influence of process parameters (rotational speed, feed rate, drill diameter and vibrational amplitude) on change in the temperature was studied using design of experiment technique i.e., response surface methodology (RSM) and data analysis was carried out using analysis of variance (ANOVA). Temperature was recorded and measured by using embedded thermocouple technique at a distance of 0.5 mm, 1.0 mm, 1.5 mm and 2.0 mm from the drill site. Statistical model was developed to predict the maximum temperature at the drill tool and bone interface. It was observed that temperature increased with increase in the rotational speed, feed rate and drill diameter and decreased with increase in the vibrational amplitude. (C) 2016 IPEM. Published by Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据