Evaluation of four automated clinical analyzers for the determination of total 25(OH)D in comparison to a certified LC-MS/MS

期刊

CLINICAL CHEMISTRY AND LABORATORY MEDICINE
卷 61, 期 8, 页码 1420-1427

出版社

WALTER DE GRUYTER GMBH
DOI: 10.1515/cclm-2022-1129

关键词

25(OH)D; CDC; immunoassay; LC-MS/MS; Milan model 1; Vitamin D Standardization-Certification Program (VDSCP); vitamin D

向作者/读者索取更多资源

The aim of this study was to compare the results of five methods for the determination of total 25(OH)D. The Cascadion method showed the highest level of concordance at medical decision cut-offs for total 25(OH)D and reached the desirable specification for bias. The presence of 3-epi-25(OH)D-3 in enriched samples was only problematic in immunoassay methods, especially Roche Elecsys methods.
Objectives The aim of this study was to compare the results of five methods for the determination of total 25(OH)D. For that purpose, two mass spectrometry and three immunoassay methods were used.Methods A total of 124 serum samples were analyzed on five different methods (i.e., a reference LC-MS/MS, Cascadion, Lumipulse, Roche Elecsys II and Roche Elecsys III). Analytical performance against LC-MS/MS was evaluated and compared to the Milan models 1 (analytical performance based on the clinical outcome using thresholds of 12, 20 and 30 ng/mL) and 2 (analytical performance based on biological variation). Additionally, imprecision studies and accuracy using NIST SRM972a samples were carried out.Results Compared to the reference LC-MS/MS method, the Lumipulse and the Roche Elecsys III assays reached the optimal criterion for bias, while the Cascadion met the desirable one. The Roche Elecsys II was not able to reach the minimal criteria. The proportion of correctly classified patients was higher using the Cascadion (95.2%) compared to the three immunoassays. In addition to its better precision, the Cascadion was not impacted by a high concentration of 3-epi-25(OH)D-3 compared to the three immunoassays.Conclusions Compared to the LC-MS/MS reference method, the Cascadion presented the highest level of concordance at medical decision cut-offs for total 25(OH)D and reached the desirable specification for bias. Moreover, the presence of 3-epi-25(OH)D-3 in enriched samples was only problematic in immunoassay methods, and especially considering Roche Elecsys methods. The release of performant fully automated mass spectrometry assays with high throughput might therefore facilitate the wide scale adoption of LC-MS/MS, even in non-specialized clinical laboratories.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据