Retrospective analysis of intra-patient laboratory variation demonstrates that the BD Vacutainer? BarricorTM blood collection tube reduces troponin variation

期刊

CLINICAL BIOCHEMISTRY
卷 114, 期 -, 页码 24-29

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2023.01.009

关键词

Preanalytic variation; Biologic variation; Analytic variation; Cardiac troponin; Electrolytes; Barricor TM; PSTTM

向作者/读者索取更多资源

This study demonstrates that the use of BD Vacutainer BarricorTM plasma blood collection tube reduces pre-analytical errors compared to the use of PSTTM tubes.
Objective: The BD Vacutainer (R) BarricorTM plasma blood collection tube uses a mechanical separator during centrifugation to separate plasma from the cellular elements of blood. Compared to use of plasma separator tubes (PSTTM) with gel, BarricorTM produces a cleaner sample with less residual cellular content. We sought to determine if BarricorTM reduces pre-analytical error compared to PSTTM.Design & Methods: We used a model previously published that utilizes serial differences between intra-patient measurements transformed into a Taylor series of variation vs time with the y-intercept equal to the sum of short-term analytic variation, preanalytic variation and biologic variation. The intra-patient variation of chlo-ride, sodium, potassium, and troponin-T (hs-TnT) obtained from the Emergency Department of a large tertiary care center sampled with PSTTM (May 2015-April 2018, n = 59,762 specimens) or BarricorTM (May 2018-May 2021, n = 61,512 specimens) was evaluated. All specimens were analyzed on either Roche Modular or Cobas (R) instruments. For each analyte, pairs of intra-patient results were tabulated and separated by 1 h in-tervals. The average between-pair variations were then regressed against time. We also determined the number of intra-patient outliers using the reference change value for each analyte.Results: The BarricorTM hs-TnT y-intercept (-0.0132) was significantly lower than the PSTTM intercept (0.9109; p = 0.022). This was also true for chloride (y-intercept = 1.0067 in BarricorTM and 1.3431 in PSTTM, p = 0.037). The percentage of hs-TnT outliers was significantly lower in BarricorTM (8.32 %) vs PSTTM (12.2 %; p < 0.001).Conclusion: The analytical and biological variations are assumed to be steady over the study periods; we ascribe the difference in the y-intercept to the preanalytical effect of the BarricorTM tube reducing platelets and other cellular debris.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据