4.4 Article

Clinical Results and Mechanical Properties of the Carotid CGUARD Double-Layered Embolic Prevention Stent

期刊

JOURNAL OF ENDOVASCULAR THERAPY
卷 24, 期 1, 页码 130-137

出版社

SAGE PUBLICATIONS INC
DOI: 10.1177/1526602816671134

关键词

carotid artery stent; closed-cell design; double-layer stent; embolic filter; internal carotid artery; in vitro testing; mechanical behavior; nitinol; open-cell design; radial force; stenosis; stent

向作者/读者索取更多资源

Purpose: To report early clinical outcomes with a novel double-layer stent for the internal carotid artery (ICA) and the in vitro investigation of the stent's mechanical properties. Methods: A prospective single-center study enrolled 30 consecutive patients (mean age 73.1 +/- 6.3 years; 21 men) with symptomatic (n=25) or high-grade (n=5) ICA stenosis treated with the new double-layer carotid CGUARD Embolic Prevention System (EPS) stent, which has an inner open-cell nitinol design with an outer closed-cell polyethylene terephthalate layer. The average stenosis of the treated arteries was 84.1%+/- 7.9% with a mean lesion length of 16.6 +/- 2.1 mm. In the laboratory, 8x40-mm stents where tested in vitro with respect to their radial force during expansion, the bending stiffness of the stent system and the expanded stent, as well as the collapse pressure in a thin and flexible sheath. The wall adaptation was assessed using fluoroscopy after stent release in step and curved vessel models. Results: The stent was successfully implanted in all patients. No peri- or postprocedural complications occurred; no minor or major stroke was observed in the 6-month follow-up. The bending stiffness of the expanded stent was 63.1 Nmm(2) and (not unexpectedly) was clearly lower than that of the stent system (601.5 Nmm(2)). The normalized radial force during expansion of the stent to 7.0 mm, consistent with in vivo sizing, was relatively high (0.056 N/mm), which correlates well with the collapse pressure of 0.17 bars. Vessel wall adaptation was harmonic and caused no straightening of the vessel after clinical application. Conclusion: Because of its structure, the novel CGUARD EPS stent is characterized by a high flexibility combined with a high radial force and very good plaque coverage. These first clinical results demonstrate a very safe implantation behavior without any stroke up to 6 months after the procedure.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据