4.5 Article

A prospective, randomized, controlled trial of robot-assisted vs freehand pedicle screw fixation in spine surgery

出版社

WILEY
DOI: 10.1002/rcs.1779

关键词

free-hand technique; posterior lumbar interbody fusion; proximal facet joint violation; randomized controlled trial; robot-assisted pedicle screw fixation

类别

资金

  1. Mazor Robotics
  2. Medtronic Inc.

向作者/读者索取更多资源

BackgroundThe purpose of this study was to compare the accuracy and safety of an instrumented posterior lumbar interbody fusion (PLIF) using a robot-assisted minimally invasive (Robot-PLIF) or a conventional open approach (Freehand-PLIF). MethodsPatients undergoing an instrumented PLIF were randomly assigned to be treated using a Robot-PLIF (37 patients) and a Freehand-PLIF (41 patients). ResultsFor intrapedicular accuracy, there was no significant difference between the groups (P = 0.534). For proximal facet joint accuracy, none of the 74 screws in the Robot-PLIF group violated the proximal facet joint, while 13 of 82 in the Freehand-PLIF group violated the proximal facet joint (P < 0.001). The average distance of the screws from the facets was 5.2 2.1 mm and 2.7 +/- 1.6 mm in the Robot-PLIF and Freehand-PLIF groups, respectively (P < 0.001). ConclusionRobotic-assisted pedicle screw placement was associated with fewer proximal facet joint violations and better convergence orientations.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据