3-D Transcranial Microbubble Cavitation Localization by Four Sensors

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TUFFC.2021.3091950

关键词

Sensors; Location awareness; Three-dimensional displays; Ultrasonic imaging; Transducers; Electron tubes; Imaging; Cavitation; focused ultrasound (FUS); microbubbles; transcranial localization

资金

  1. Office of Naval Research [N00014-19-1-2335]
  2. National Institutes of Health [R01EB027223, R01EB030102, R01MH116981]

向作者/读者索取更多资源

This study investigates the feasibility of using four sensors for transcranial 3-D localization of cavitation during focused ultrasound therapy in the brain. The proposed method achieved accurate cavitation localization when the source was close to the geometric center of the sensor network. The simplicity and accuracy of this approach make it a promising technique for transcranial cavitation detection.
Cavitation is the fundamental physical mechanism of various focused ultrasound (FUS)-mediated therapies in the brain. Accurately knowing the three-dimensional (3-D) location of cavitation in real-time can improve the targeting accuracy and avoid off-target tissue damage. Existing techniques for 3-D passive transcranial cavitation detection require the use of expensive and complicated hemispherical phased arrays with 128 or 256 elements. The objective of this study was to investigate the feasibility of using four sensors for transcranial 3-D localization of cavitation. Differential microbubble cavitation detection combined with the time difference of arrival algorithm was developed for the localization using the four sensors. Numerical simulation using k-Wave toolbox was performed to validate the proposed method for transcranial cavitation source localization. The sensors with a center frequency of 2.25 MHz and a 6 dB bandwidth of 1.39 MHz were used to locate cavitation generated by FUS (500 kHz) sonication of microbubbles that were injected into a tube positioned inside an ex vivo human skullcap. Cavitation emissions from the microbubbles were detected transcranially using the four sensors. Both simulation and experimental studies found that the proposed method achieved accurate 3-D cavitation localization. When the cavitation source was located within 30 mm from the geometric center of the sensor network, the accuracy of the localization method with the skull was measured to be 1.9 +/- 1.0 mm, which was not significantly different from that without the skull (1.7 +/- 0.5 mm). The accuracy decreased as the cavitation source was away from the geometric center of the sensor network. It also decreased as the pulse length increased. Its accuracy was not significantly affected by the sensor position relative to the skull. In summary, four sensors combined with the proposed localization algorithm offer a simple approach for 3-D transcranial cavitation localization.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据