Safety and chronic lesion characterization of pulsed field ablation in a Porcine model

期刊

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY
卷 32, 期 4, 页码 958-969

出版社

WILEY
DOI: 10.1111/jce.14980

关键词

atrial fibrillation; catheter ablation; electroporation; pulsed electric field modeling; pulsed field ablation

资金

  1. Medtronic Inc.

向作者/读者索取更多资源

Pulsed field ablation (PFA) as an alternative to thermal-based ablation systems for treating atrial fibrillation patients has been studied in a porcine model. The chronic effects and safety of overlapping lesions created by the PFA system were evaluated, showing complete circumferential replacement fibrosis at 4-week postablation with no significant adverse events. The study demonstrated the non-thermal nature of PFA and its excellent safety profile for chronic myocardial and collateral tissue.
Background Pulsed field ablation (PFA) has been identified as an alternative to thermal-based ablation systems for treatment of atrial fibrillation patients. The objective of this Good Laboratory Practice (GLP) study was to characterize the chronic effects and safety of overlapping lesions created by a PFA system at intracardiac locations in a porcine model. Methods A circular catheter with nine gold electrodes was used for overlapping low- or high-dose PFA deliveries in the superior vena cava (SVC), right atrial appendage (RAA), and right superior pulmonary vein (RSPV) in six pigs. Electrical isolation was evaluated acutely and chronic lesions were assessed via necropsy and histopathology after 4-week survival. Acute and chronic safety data were recorded peri- and post-procedurally. Results No animal experienced ventricular arrhythmia during PFA delivery, and there was no evidence of periprocedural PFA-related adverse events. Lesions created in all anatomies resulted in electrical isolation postprocedure. Lesions were circumferential, contiguous, and transmural, with all converting into consistent lines of chronic replacement fibrosis, regardless of trabeculated or smooth endocardial surface structure. Ablations were non-thermally generated with only minimal post-delivery temperature rises recorded at the electrodes. There was no evidence of extracardiac damage, stenosis, aneurysms, endocardial disruption, or thrombus. Conclusion PFA deliveries to the SVC, RAA, and RSPV resulted in complete circumferential replacement fibrosis at 4-week postablation with an excellent chronic myocardial and collateral tissue safety profile. This GLP study evaluated the safety and efficacy of a dosage range in preparation for a clinical trial and characterized the non-thermal nature of PFA.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据