Postmarket Surveillance of Point-of-Care Glucose Meters through Analysis of Electronic Medical Records

期刊

CLINICAL CHEMISTRY
卷 62, 期 5, 页码 716-724

出版社

AMER ASSOC CLINICAL CHEMISTRY
DOI: 10.1373/clinchem.2015.251827

关键词

-

资金

  1. NIH

向作者/读者索取更多资源

BACKGROUND: The electronic medical record (EMR) holds a promising source of data for active postmarket surveillance of diagnostic accuracy, particularly for point-of-care (POC) devices. Through a comparison with prospective bedside and laboratory accuracy studies, we demonstrate the validity of active surveillance via an EMR data mining method [Data Mining EMRs to Evaluate Coincident Testing (DETECT)], comparing POC glucose results to near-in-time central laboratory glucose results. METHODS: The Roche ACCU-CHEK Inform II (R) POC glucose meter was evaluated in a laboratory validation study (n = 73), a prospective bedside intensive care unit (ICU) study (n = 124), and with DETECT (n = 852-27 503). For DETECT, the EMR was queried for POC and central laboratory glucose results with filtering based on of bedside collection timestamps, central laboratory time delays, patient location, time period, absence of repeat testing, and presence of peripheral lines. RESULTS: DETECT and the bedside ICU study produced similar estimates of average bias (4.5 vs 5.0 mg/dL) and relative random error (6.3% vs 5.6%), with overlapping CIs. For glucose <100 mg/dL, the laboratory validation study estimated a lower relative random error of 3.6%. POC average bias correlated with central laboratory turnaround times, consistent with 4.8 mg . dL(-1) . h(-1) glycolysis. After glycolysis adjustment, average bias was estimated by the bedside ICU study at -0.4 mg/dL (CI, -1.6 to 0.9) and DETECT at -0.7 (CI, -1.3 to 0.2), and percentage POC results occurring outside Clinical Laboratory Standards Institute quality goals were 2.4% and 4.8%, respectively. CONCLUSIONS: This study validates DETECT for estimating POC glucose meter accuracy compared with a prospective bedside ICU study and establishes it as a reliable postmarket surveillance methodology. (C) 2016 American Association for Clinical Chemistry

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据