Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: a cross-sectional method comparison study

期刊

BMC ANESTHESIOLOGY
卷 20, 期 1, 页码 -

出版社

BMC
DOI: 10.1186/s12871-020-01008-8

关键词

Agreement; Bias; Correlation; Hemoglobin measurement; Point-of-care testing

资金

  1. Faculty of Medicine Ramathibodi Hospital, Mahidol University, Thailand [RF_59037]

向作者/读者索取更多资源

Background Transfusion decision during the perioperative period mostly relies on the point-of-care testing for Hb measurement. This study aimed systematically compared four point-of-care methods with the central laboratory measurement of hemoglobin (LHb) regarding the accuracy, precision, and assay practicality to identify the preferred point-of-care method during the perioperative period. Methods This cross-sectional method comparison study was conducted in the surgical intensive care unit at Ramathibodi Hospital, Thailand, from September 2015 to July 2016. Four point-of-care methods, i.e., capillary hematocrit (HctCap), HemoCue Hb201+, iSTAT with CG8+ cartridge, and SpHb from Radical-7 pulse co-oximeter were carried out when LHb was ordered. Pearson correlation and Bland-Altman analyses were performed to assess the accuracy and precision, while the workload, turnaround time, and the unit cost were evaluated for the method practicality. Results Thirty-five patients were enrolled, corresponding to 48 blood specimens for analyses, resulting in the measured hemoglobin of 11.2 +/- 1.9 g/dL by LHb. Ranking by correlation (r), mean difference (bias) and 95% limit of agreement (LOA) showed the point-of-care methods from the greater to the less performance as followed, iSTAT-LHb pair (r = 0.941; bias 0.15 (95% LOA; - 1.41, 1.12) g/dL), HemoCue-LHb pair (r = 0.922; bias - 0.18 (95% LOA; - 1.63, 1.28) g/dL), SpHb-LHb pair (r = 0.670; bias 0.13 (95% LOA; - 3.12, 3.39) g/dL) and HctCap-LHb pair (r = 0.905; bias 0.46 (95% LOA; - 1.16, 2.08) g/dL). Considering the practicality, all point-of-care methods had less workload and turnaround time than LHb, but only HemoCue and HctCap had lower unit cost. Conclusion This study identified HemoCue as the suitable point-of-care method for the sole purpose of Hb measurement in the surgical ICU setting, while iSTAT should be considered when additional data is needed.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据