期刊
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY
卷 31, 期 8, 页码 2136-2147出版社
WILEY
DOI: 10.1111/jce.14414
关键词
atrial fibrillation; catheter ablation; pulsed-field ablation
Pulsed-field ablation (PFA) is a promising new ablation modality for the treatment of atrial fibrillation. This energy form employs a train of microsecond duration high amplitude electrical pulses that ablate myocardium by electroporation of the sarcolemmal membrane without measurable tissue heating. The ablation pulse waveform has multiple variable components that can affect ablation efficacy, thus each proprietary system has unique properties that cannot be generalized to other systems. Success with PFA depends upon the proximity of the electrode to the target tissue, but not necessarily upon contact. A unique feature of PFA is tissue specificity. Myocardium is very susceptible to irreversible injury whereas the esophagus, phrenic nerves, pulmonary veins, and coronary arteries are relatively resistant to injury. The tissue specificity of PFA may result in a wide therapeutic range and improved safety profile during atrial fibrillation ablation. Vein isolation can be achieved very rapidly (seconds) promising that PFA may reduce procedure time to 1 hour or less. This attractive new technology promises to be a major advance in the field of atrial fibrillation ablation.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。