Electroosmotic flow driven microfluidic device for bacteria isolation using magnetic microbeads

期刊

SCIENTIFIC REPORTS
卷 9, 期 -, 页码 -

出版社

NATURE PUBLISHING GROUP
DOI: 10.1038/s41598-019-50713-z

关键词

-

资金

  1. National Institute for Occupational Safety and Health through the University of Cincinnati Education and Research Center Grant [T42OH008432]

向作者/读者索取更多资源

The presence of bacterial pathogens in water can lead to severe complications such as infection and food poisoning. This research proposes a point-of-care electroosmotic flow driven microfluidic device for rapid isolation and detection of E. coli in buffered solution (phosphate buffered saline solution). Fluorescent E. coli bound to magnetic microbeads were driven through the microfluidic device using both constant forward flow and periodic flow switching at concentrations ranging from 2 x 10(5) to 4 x 10(7) bacteria/mL. A calibration curve of fluorescent intensity as a function of bacteria concentration was created using both constant and switching flow, showing an increase in captured fluorescent pixel count as concentration increases. In addition, the use of the flow switching resulted in a significant increase in the capture efficiency of E. coli, with capture efficiencies up to 83% +/- 8% as compared to the constant flow capture efficiencies (up to 39% +/- 11%), with a sample size of 3 mu L. These results demonstrate the improved performance associated with the use of the electroosmotic flow switching system in a point-of-care bacterial detection assay.