On the Equivalence of FCS and FRAP: Simultaneous Lipid Membrane Measurements

期刊

BIOPHYSICAL JOURNAL
卷 111, 期 1, 页码 152-161

出版社

CELL PRESS
DOI: 10.1016/j.bpj.2016.06.001

关键词

-

资金

  1. Singapore Ministry of Education [MOE2012-T2-1-101]
  2. Research Centre of Excellence Grant from the Ministry of Education, Singapore [VA 5IOBX000372]

向作者/读者索取更多资源

Fluorescence correlation spectroscopy (FCS) and fluorescence recovery after photobleaching (FRAP) are widely used methods to determine diffusion coefficients. However, they often do not yield the same results. With the advent of camera-based imaging FCS, which measures the diffusion coefficient in each pixel of an image, and proper bleaching corrections, it is now possible to measure the diffusion coefficient by FRAP and FCS in the exact same images. We thus performed simultaneous FCS and FRAP measurements on supported lipid bilayers and live cell membranes to test how far the two methods differ in their results and whether the methodological differences, in particular the high bleach intensity in FRAP, the bleach corrections, and the fitting procedures in the two methods explain observed differences. Overall, we find that the FRAP bleach intensity does not measurably influence the diffusion in the samples, but that bleach correction and fitting introduce large uncertainties in FRAP. We confirm our results by simulations.