Electrical Impedance Tomography for Cardio-Pulmonary Monitoring

期刊

JOURNAL OF CLINICAL MEDICINE
卷 8, 期 8, 页码 -

出版社

MDPI
DOI: 10.3390/jcm8081176

关键词

electrical impedance tomography; bioimpedance; image reconstruction; thorax; regional ventilation; regional perfusion; monitoring

资金

  1. German Research Council (Deutsche Forschungsgemeinschaft, DFG) [PU 219/2-1]
  2. Department of Anesthesiology and Intensive Care Medicine, University Hospital Bonn, Germany

向作者/读者索取更多资源

Electrical impedance tomography (EIT) is a bedside monitoring tool that noninvasively visualizes local ventilation and arguably lung perfusion distribution. This article reviews and discusses both methodological and clinical aspects of thoracic EIT. Initially, investigators addressed the validation of EIT to measure regional ventilation. Current studies focus mainly on its clinical applications to quantify lung collapse, tidal recruitment, and lung overdistension to titrate positive end-expiratory pressure (PEEP) and tidal volume. In addition, EIT may help to detect pneumothorax. Recent studies evaluated EIT as a tool to measure regional lung perfusion. Indicator-free EIT measurements might be sufficient to continuously measure cardiac stroke volume. The use of a contrast agent such as saline might be required to assess regional lung perfusion. As a result, EIT-based monitoring of regional ventilation and lung perfusion may visualize local ventilation and perfusion matching, which can be helpful in the treatment of patients with acute respiratory distress syndrome (ARDS).