期刊
PHARMACEUTICAL RESEARCH
卷 36, 期 11, 页码 -出版社
SPRINGER/PLENUM PUBLISHERS
DOI: 10.1007/s11095-019-2690-8
关键词
monoclonal antibody (mAb); protein aggregation; analytical characterization; size exclusion chromatography (SEC); dynamic light scattering (DLS); micro flow imaging (MFI); sub-visible particles
资金
- Centre of Excellence for Biopharmaceutical Technology grant from Department of Biotechnology, Government of India [BT/COE/34/SP15097/2015]
Purpose To develop an analytical platform for the estimation as well as characterization of aggregates over the complete size spectrum (from invisible monomer to visible precipitates). Methods Two mAb samples were incubated at 30 degrees C in different buffer systems of protein A chromatography for observing degradation due to aggregation. The aggregation in these samples was quantified by size exclusion chromatography (SEC), dynamic light scattering (DLS), and micro flow imaging (MFI). Results The results obtained from various characterization tools were analysed in various size ranges - size exclusion chromatography (SEC) (1 nm - 25 nm), dynamic light scattering (DLS) (10 nm - 5 mu m), and micro flow imaging (MFI) (2 mu m - 300 mu m). Since each characterization tool covers a particular size range, data from multiple tools was collected in the handover regions to demonstrate accuracy of the platform. Conclusions Based on the observations from the experiments, an analytical platform has been proposed covering the whole size spectrum that would be of utility to those engaged in formulation development as well as other aspects related to stability of biotherapeutic products.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据