Quantification of human bone microarchitecture damage in press-fit femoral knee implantation using HR-pQCT and digital volume correlation

出版社

ELSEVIER
DOI: 10.1016/j.jmbbm.2019.04.054

关键词

Press-fit implantation; Uncemented total knee replacement; Trabecular bone deformation; HR-pQCT; Digital volume correlation

资金

  1. DePuy Synthes Joint Reconstruction (Leeds, UK)

向作者/读者索取更多资源

Primary press-fit fixation of femoral knee prostheses is obtained thanks to the inside dimensions of the implant being undersized with respect to the bone cuts created intra-operatively, dictated by a press-fit specified by the implant design. However, during prostheses press-fit implantation, high compressive and shear stresses at the implant-bone interface are generated, which causes permanent bone damage. The extent of this damage is unknown, but it may influence the implant stability and be a contributing factor to aseptic loosening, a main cause of revisions for knee arthroplasty. The aim of this ex-vivo study was to quantify, using high-resolution peripheral quantitative computed tomography (HR-pQCT) imaging and Digital Volume Correlation (DVC), permanent bone deformation due to press-fit femoral knee implantation of a commonly used implant. Six human cadaveric distal femora were resected and imaged with HR-pQCT (60.7 mu m/voxel, isotropic). Femurs were fitted with cementless femoral knee implants (Sigma PFC) and rescanned after implant removal. For each femur, permanent deformation was examined in the anterior, posterior-medial and posterior-lateral condyles for volumes of interest (VOIs) of 10 mm depth. The bone volume fraction (BV/TV) for the VOIs in pre- and post-implantation images was calculated, at increasing depth from the bone surface. DVC was applied on the VOIs pre- and post-implantation, to assess trabecular bone displacements and plastically accumulated strains. The BV/TVpost/BV/TVpre ratio vs. depth showed, consistently among the six femurs, three consecutive points of interest at increasing bone depth, indicating: bone removal (ratio < 100%), compaction (ratio > 100%) and no changes (ratio = 100%). Accordingly, the trabecular bone displacement computed by DVC suggested bone compaction up to 2.6 +/- 0.8 mm in depth, with peak third principal strains of -162,100 +/- 55,000 mu epsilon (mean absolute error: 1,000-2,000 mu epsilon, SD: 200-500 mu epsilon), well above the yield strain of bone (7,000-10,000 mu epsilon). Combining 3D-imaging, at spatial resolutions obtainable with clinical HR-pQCT, and DVC, determines the extent of plastic deformation and accumulated compressive strains occurring within the bone due to femoral press-fit implantation. The methods and data presented can be used to compare different implants, implant surface coatings and press-fit values. These can also be used to advance and validate computational models by providing information about the bone-implant interface obtained experimentally. Future studies using these methods can assist in determining the influence of bone damage on implant stability and the subsequent osseointegration.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据