Direct determination of glyphosate, glufosinate, and AMPA in soybean and corn by liquid chromatography/tandem mass spectrometry

期刊

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 408, 期 18, 页码 4995-5004

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-016-9597-6

关键词

Glyphosate; Soybean; Corn; LC-MS/MS; Direct determination

资金

  1. U.S. Food and Drug Administration

向作者/读者索取更多资源

Glyphosate, glufosinate, and aminomethylphosphonic acid (AMPA) are amphoteric, low mass, high water soluble, and do not have chromophore. They are very difficult to be retained on a reversed phase HPLC and detected by UV or fluorescence detectors. A liquid chromatography/tandem mass spectrometry (LC-MS/MS) method was developed to determine these analytes in soybean and corn using a reversed phase with weak anion-exchange and cation-exchange mixed-mode Acclaim (TM) Trinity (TM) Q1 column. The sample was shaken with water containing ethylenediaminetetraacetic acid disodium salt (Na(2)EDTA) and acetic acid for 10 min to precipitate protein and extract the analytes into the solution. The supernatant was passed thru an Oasis HLB SPE to retain suspended particulates and non-polar interferences. The sample was directly injected and analyzed in 6 min by LC-MS/MS with no sample concentration or derivatization steps. Three isotopically labeled internal standards corresponding to each analyte were used to counter matrix suppression effect. Linearity of the detector response with a minimum coefficient of determination (R (2)) of more than 0.995 was demonstrated in the range of 10 to 1000 ng/mL for each analyte. Accuracy (recovery %) and precision (relative standard deviation or RSD %) were evaluated at the fortification levels of 0.1, 0.5, and 2 mu g/g in seven replicates in both soybean and corn samples.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据