4.6 Article

An Optimized Colorimetric Readout Method for Lateral Flow Immunoassays

期刊

SENSORS
卷 18, 期 12, 页码 -

出版社

MDPI
DOI: 10.3390/s18124084

关键词

colorimetry; photometry; assay reader; quantification

资金

  1. National Research Foundation of Korea (NRF) - Ministry of Science and ICT [NRF-2016R1A2B1011869]
  2. National Research Foundation of Korea [2016R1A2B1011869] Funding Source: Korea Institute of Science & Technology Information (KISTI), National Science & Technology Information Service (NTIS)

向作者/读者索取更多资源

Despite its broad penetration of various markets, the quantitative lateral flow immunoassay (LFIA) suffers from sensitivity issues in some cases. To solve this problem, an optimized colorimetric readout method for LFIA quantification is proposed in this study. An assay reader device utilizing a color camera and an analysis method using a Bayer filtered image were developed. Spectrometric measurements of the assay test line were performed to determine the color channel that contains the test line information and effectively minimizes noise. The change in the intensity ratio with increasing concentration of the target substance in the sample was largest in the green channel. The linear range of the output curve ranged from 0 to 10 ng/mL, and the detection limit was 2 ng/mL. The suggested instrumentation and analysis methods are expected to effectively resolve the low-sensitivity problems of the former LFIA systems and to offer other prospective functionalities for LFIA quantification.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据