4.0 Article

In Vivo Evaluation of Customized Pulse Oximeter and Sensitivity Pulp Tests for Assessment Of Pulp Vitality

期刊

JOURNAL OF CLINICAL PEDIATRIC DENTISTRY
卷 43, 期 1, 页码 11-15

出版社

JOURNAL PEDODONTICS INC
DOI: 10.17796/1053-4625-43.1.3

关键词

Pulp vitality; Pulp test; Pulse oximeter

向作者/读者索取更多资源

Objective: The purpose of this study was to lest a customized pulse oximeter (CPO) for evaluation of pulp vitality in primary and permanent teeth against clinical diagnosis (vital and untreated non-vital) in order to expand its clinical use for pulp preservation. Study design: CPO was evaluated on intact primary and permanent central or lateral incisor (CI, LI) teeth- vital (group 1, 20n each); untreated non-vital (group 2, 10n each) and; root filled non-vital (group 3, 10n each) of children 4-12 years according to inclusion/exclusion criteria. For each patient CPO was first applied on finger followed by vitality tests in following sequence- electrical, pulse oximetry and thermal tests. Results: Mean oxygen saturation (%SpO2) in permanent and primary-vital teeth was 88.78% & 87.77% respectively; non-vital teeth was 74.67% & 75.00% respectively; and in all root filled teeth was 0%. Tooth and finger oxygen saturation values showed strong positive relationship in vital primary or permanent teeth and; no correlation in untreated non-vital primary or permanent teeth. The accuracy rate of thermal pulp test and pulse oximetry was 100% and for electrical pulp test it was 90% for permanent and 86.67% for primary teeth. Conclusion: The CPO tested in this study proved to be a valuable adjunct for diagnosing pulp vitality by objective means.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据